Material Compatibilities; Disinfecting Surfaces; Cleaning Surfaces; Cleaning Handle In An Ultrasonic Bath 31 Sterilising Handle In An Autoclave - Cheyenne HAWK PEN Operating Instructions Manual

+ hawk pen tray
Hide thumbs Also See for HAWK PEN:
Table of Contents

Advertisement

7.1

Material compatibilities

Wherever possible, use one of the
following disinfectants
Manufac-
Product
turer
Big Spray
Antiseptica
'new'
Bode
Bacillol
Chemie
Incidin
Ecolab
Foam
Schülke &
Mikrozid
Mayr
Liquid
Based on the following material compatibilities,
you can assess which other cleaning agents
and disinfectants are suitable
The anodised aluminium components of
the HAWK PEN are resistant to:
Weak acids (e.g. acetic acid ≤ 10%, citric
acid ≤ 10%)
Aliphatic fluorocarbons (e.g. pentane,
hexane)
Ethanol
Most inorganic salts and their aqueous
solutions (e.g. sodium chloride, calcium
chloride, magnesium sulphate)
31 EN
The anodised aluminium components of
the HAWK PEN are not resistant to:
Strong acids (e.g. hydrochloric acid,
sulphuric acid, nitric acid)
Oxidising acids (e.g. peracetic acid)
Exposure time
Lyes (e.g. caustic soda, ammonia and
all substances with a pH value > 7)
1 to 5 mins
Aromatic/halogenised
(e.g. phenol, chloroform) as well as
30 s to 1 min
acetone and benzene
1 to 2 mins
7.2

Disinfecting surfaces

Before and after each usage:
1 to 2 mins
Disinfect all surfaces of the HAWK PEN
and its accessories as described in
chapter "Disinfecting equipment" on
page 25.
7.3

Cleaning surfaces

In the event of external contamination:
Wipe the HAWK PEN and the protective
hose with a soft cloth dipped in a
cleaning agent or a disinfectant.
Clean inaccessible surfaces, e.g.the
thread of the drive unit, with a soft-
bristled nylon brush.
7.4
Cleaning handle in an ultrasonic
bath
If the handle is severely soiled:
Use cleaning agent Tickomed 1 or
Tickopur
Dr.  H.  Stamm GmbH in accordance
with the manufacturer's instructions
in a concentration of 3% in cold water.
fluorocarbons
Conduct the cleaning operation for
10 minutes at a frequency of 35 kHz.
Dry the handle fully after cleaning,
before you use it again.
7.5
Sterilising handle in an
autoclave
Whenever handle is contaminated:
Perform all operations described in
chapter 'Disinfecting equipment' (see
Page 25).
Autoclave the handle for 20 minutes at
a temperature of 121°C and a pressure
of 2 bar.
Dry the handle fully after cleaning,
before you use it again.
R33
from
manufacturer
Cheyenne HAWK PEN