7. Maintenance cont.
7-1 To Insert the Bur
3) Rotate the handpiece and immerse half of the handpiece head in the cup
of water. (Fig. 5)
4) Rotate then stop intermittently the handpiece 3 times for 2 to 3 seconds
each time.
5) Wipe the handpiece dry.
7-2 Cleaning (Handpiece)
1) Remove dirt and debris from the handpiece. Do not use a wire brush.
2) Wipe clean with alcohol-immersed cotton swab or cloth.
This icon denotes that the handpiece can be washed via Thermo-Disinfector.
Refer to the Thermo-Disinfector manual
.
Fig. 5