National Instruments 446 Series Calibration Procedure page 22

Table of Contents

Advertisement

NI-DAQmx Function Call
Call
DAQmxCfgSampClkTiming
the following parameters:
taskHandle:
myTaskHandle
source:
NULL
rate:
204800.0
activeEdge:
DAQmx_Val_Rising
SampleMode:
DAQmx_Val_ContSamps
sampsPerChanToAcquire:
NI-DAQmx Function Call
Call
DAQmxWriteAnalogF64
following parameters:
taskHandle:
myTaskHandle
numSampsPerChan:
autoStart:
FALSE
timeout:
10.0
dataLayout:
DAQmx_Val_GroupByChannel
writeArray:
myDataArray
reserved:
NULL
sampsPerChanWritten:
int32PointerReturnValue
NI 446X Calibration Procedure
7.
Configure the timing properties of the generation using the following
function.
with
1024
8.
Configure the onboard memory of the device using the following
function and the array created in step 2.
with the
1024
LabVIEW Block Diagram
LabVIEW Block Diagram
22
ni.com

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pci-4461Pxi-4461Pci-4462Pxi-4462

Table of Contents