National Instruments 446 Series Calibration Procedure page 17

Table of Contents

Advertisement

NI-DAQmx Function Call
Call
DAQmxCfgSampClkTiming
following parameters:
taskHandle:
myTaskHandle
source:
NULL
rate:
204800.0
activeEdge:
DAQmx_Val_Rising
sampleMode:
DAQmx_Val_FiniteSamps
sampsPerChanToAcquire:
NI-DAQmx Function Call
Call
DAQmxStartTask
parameters:
taskHandle:
myTaskHandle
© National Instruments Corporation
6.
Configure the timing properties of the acquisition using the following
function.
with the
102400
7.
Start the task using the following function.
with the following
LabVIEW Block Diagram
LabVIEW Block Diagram
17
NI 446X Calibration Procedure

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pci-4461Pxi-4461Pci-4462Pxi-4462