6 MAINTENANCE & CARE
For accurate readings keep your fingers away from, and avoid obstructing airflow through, the holes in the
casing around the humidity and air temperature sensors. Some time (~20 minutes) may be required to
acclimatise the gauge once it is brought from a cold into a warm environment or vice versa
Regular calibration checks over the life of the gauge are a requirement of ISO 9000 and other quality
procedures. For certification please contact Elcometer or your supplier.
The instrument does not contain any user-serviceable components. In the unlikely event of a fault, the
instrument should be returned to your local Elcometer supplier or direct to Elcometer. The warranty will be
invalidated if the instrument has been opened.
To avoid damage:
When measuring surface temperature a light contact between the probe and the surface is sufficient -
DO NOT USE EXCESSIVE FORCE
Do not drag the surface temperature probe over the surface
Keep the gauge away from paint dust and dirt as much as possible as these can affect the long term
accuracy of the humidity sensor
The gauge is dustproof and resistant to water jets and should not be affected by dust and moisture
under normal conditions, however, the gauge should not be submerged in liquids