Choose Flash Image Window; Autosampler Normal Configuration; Verification Of Firmware Upgrade - CETAC ASX-520 Upgrading And Servicing Manual

Autosamplers with an exr-8
Hide thumbs Also See for ASX-520:
Table of Contents

Advertisement

Guide for Upgrading the
Figure 1-16. Choose Flash Image Window.
3. Select the OK button. A progress bar will appear.
4. When it is complete, the Rabbit Field Utility can be closed.

Autosampler Normal Configuration

1. Turn off the autosampler.
2. On the board move the jumper from J5 back to J3 (Figure 1-15).
3. Turn the autosampler back on. The autosampler should go to the home position.

Verification of Firmware Upgrade

1. Start HyperTerminal.
Operating a CETAC Autosampler using HyperTerminal).
2. In HyperTerminal type VERSS. The system should respond with the new firmware version.
3. In HyperTerminal type SETTYPE=nnn where nnn is the configuration number.
Configuration
Number
23
24
25
26
Firmware on the ASX-520 and ASX-520HS Autosamplers with an EXR-8
(For instructions on using HyperTerminal, refer to the Guide to
Configuration Type
CETAC Standard
CETAC Standard, High Speed
PE
PE High Speed