Unit description
Protokoll:
Example::
15.05.01
YAEE0013
− The interface operates with 1 stop bit, no parity bit and 8 data bits.
− Transfer rate either 2400, 4800, 9600 (factory setting) or 19200 baud as selected.
− The RS232 interface can be operated with or without hardware handshake (RTS/CTS).
− The command from the computer must be terminated with CR, CRLF or LFCR.
− The response of the thermostat is always terminated with CRLF.
CR = carriage return (hex: 0D)
LF = line feed (hex: 0A)
Transfer of setpoint 30.5°C to the thermostat.
Computer
„OUT_SP_00_30.5"CRLF
Thermostat
"
#
„OK"CRLF
7