Rs 485 Interface - Lauda Ecoline E 200 Operating Instructions Manual

Immersion thermostats; bath/ circulation thermostats
Table of Contents

Advertisement

Unit description
3.5.2

RS 485 Interface

Connecting cable:
pin
1
5
6
The RS 485 bus absolutely needs a bus termination in form of a terminating network which ensures a
defined unattended time in the high-resistance phases. The bus terminal is defined as
Generally this network is integrated on the plug-in card of the computer (RS 485).
Protocol:
8
Thermostat
9-pin sub-D-socket
Data
Data A
SG (Signal Ground)
optional
Data B
− Use screened connecting cables.
− Connect the screen to the plug case.
− The lines are electrically isolated from the remaining electronics.
− Unoccupied pins must not be connected!
− The interface operates with 1 stop bit, no parity bit and 8 data bits.
− Transfer rate either 2400, 4800, 9600 (factory setting) or 19200 baud as selected.
− The RS485 commands are always preceded by the device address. There is
provision for 127 addresses. The address must always have 3 digits. (A000_... to
A127_...).
− The command from the computer must be terminated with CR.