Calibration Checks - Testing Instructions; (For Trained Specialist Staff Only) - Bosch+Sohn BoSo Medicus Smart User Instructions

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 36

Calibration checks - Testing instructions

(for trained specialist staff only)

Calibration checks – every 2 years at the
latest – must be performed by the follo-
wing institutions or persons:
– Manufacturer
EN
– Trade measurements authorities
– Persons who fulfil the requirements of
the legislation governing the operation
of medical devices.
A) Function testing
Function testing of the monitor can
only be carried out on a person or with
a suitable simulator.
B)Testing of Density of Pressure
Circuit
Pressure Indication
68
and
Divergence
1. Remove battery.
2. Hold Start button
the same time insert battery.
3. Release Start button
is ready for testing. The displays of SYS
and DIA now show the actual pressu-
re.
4. Prior to testing the density, open the
battery compartment and take the T-
connector off the unit. Wrap the cuff
around a fixed cylinder and connect it
with the T-connector (without deflati-
from
on valve) to the pump ball and the
unit.
pressed and at
. The unit now

Advertisement

Table of Contents
loading

Table of Contents