Error Messages - Bosch+Sohn BoSo Medicus Smart User Instructions

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 36

Error messages

If any problems occur during the measurement, the display will show an error message
instead of a blood pressure reading.
Meaning of error messages:
EN
60
Cause of Err message:
Excessive variation in cuff pressure Hold arm still
during measurement
No valid pulse readings
Systole – diastole difference
too small:
Systole – diastole < _ 10 mmHg
Cause of Err CuF message:
Incorrectly inflated
Troubleshooting:
Check position of cuff (see page 53)
and repeat measurement.
Check position of cuff (see page 53)
and repeat measurement.
Troubleshooting:
Cuff possibly too loose, attach
cuff more firmly (see page 54)

Advertisement

Table of Contents
loading

Table of Contents