Calibration checks - Testing instructions
(for trained specialist staff only)
Calibration checks – every 2 years at the
latest – must be performed by the follow-
ing institutions or persons:
– Manufacturer
– Trade measurements authorities
– Persons who fulfil the requirements of
EN
the legislation governing the operation
of medical devices.
A) Function testing
Function testing of the monitor can only
be carried out on a person or with a
suitable simulator.
B) Testing of pressure circuit integrity
and deviation of pressure display
82
Note:
a) When the pressure in measurement
mode increases above 299 mmHg, the
quick release valve is triggered and the
pressure circuit opens. When the press-ure
in test mode increases above 320 mmHg,
the measured value display flashes.
b) A pump ball must also be introduced into
the pressure circuit for the tests.
Testing
1.) Remove batteries.
2.) Hold START button pressed down and
insert batteries.