Calibration checks - Testing instructions
(for trained specialist staff only)
Mind the setting time for cuff: at least
30 seconds.
Accomplish testing in the usual man-
ner.
5. Prior to testing the divergence of the
pressure indication, open the battery
compartment and take the T-connec-
tor off the unit. Connect the unit with
the normal unit and a pump ball.
Accomplish testing in the usal man-
ner.
6. Switch off unit.
7. Re-connect the cuff and the pumpball
with the T-connector.
C) Placing of Safety Stamps
For complete safety cover up the drill
hole for the base screw. Alternatively,
connect upper housing and lower
housing with a safety stamp placed
directly on one of the housing edges.
The device must not undergo service
or maintenance while it is in use.