Flow Calibration Menu - Process Technology Lufran LUF Series Manual

Fluoropolymer inline water heater
Table of Contents

Advertisement

Engineering Information (Continued):
Operator Interface Panel (OIP) (Continued):

Flow Calibration Menu:

The flow calibration menu is accessed through the secure menu only. Refer to the engineering
information section of this manual for the flow meter calibration procedure.
SLC Setup Information
Panel
HIGH FLOW
HIGH RAW
SLOPE CALC.
RAW DATA
LOW FLOW
LOW RAW
OFFSET CALC.
FLOW CALC.
SLOPE
OFFSET
BACK
M-43-25
Revision - Date 03-01-04-18
Figure 17: Flow Calibration Menu
Manually-entered flow rate used as the high-end calibration point,
taken from the display on the flow meter itself.
Manually entered raw data value that corresponds to the high flow
value, taken from the RAW DATA display on this menu.
Calculated value from the HIGH FLOW, HIGH RAW, LOW FLOW
and LOW RAW fields.
Information from the flow meter.
Manually-entered flow rate used as the low-end calibration point,
taken from the display on the flow meter itself.
Manually entered raw data value that corresponds to the low flow
value, taken from the RAW DATA display on this menu.
Calculated value from the HIGH FLOW, HIGH RAW, LOW FLOW
and LOW RAW fields.
Calculated value from the SLOPE and OFFSET fields. When the
flow meter is calibrated, this value should match the display on the
flow meter.
Manually-entered field taken from the SLOPE CALC. field.
Manually-entered field taken from the OFFSET CALC. field.
This button will return to the previous screen.
27
Description
LUF-Series Lufran Manual