R&S RT-ZC31 User Manual page 19

Current probe
Table of Contents

Advertisement

®
R&S
RT-ZC31
4. Check that the conductor being measured is not clamped when supplying
power to the R&S RT-ZC31. When power is turned on, a demagnetizing wave-
form is initially applied to the output. This is intentional in the design, and not a
fault.
5. Turn on the power switch R&S RT-ZA13 probe power supply. Ensure that the
front panel power indicator lights.
The POWER LED lights up in green and the DEMAG/AUTO ZERO LED is
slowly flashing in orange.
6. Connect the output connector of the R&S RT-ZC31 to one of the BNC input
connectors of the oscilloscope. Turn the collar until it clicks, and check that it
is locked securely.
Figure 3-1: Oscilloscope inputs
The current waveform of the measured conductor is output at a constant rate
(1 V/A).
7. Configure the probe connection at the oscilloscope. Make sure to set the fol-
lowing:
● Vertical unit = Ampere
● Coupling = DC
● Termination = 1 MΩ
● Manual Gain = 1 V/A
Alternatively, select "Predefined probe" = R&S RT-ZC31 if this selection is
available on the instrument.
The procedure depends on the used instrument and is described in the oscil-
loscope's user manual. Supported oscilloscopes are listed in the R&S RT-Zxx
data sheet.
8. When disconnecting the output connector, be sure to release the lock before
pulling off the connector. Forcibly pulling the connector without releasing the
lock, or pulling on the cable can damage the terminator.
User Manual 1801.4949.02 ─ 01
Measuring Procedure
Preparing the Measurement
19

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rt-za13

Table of Contents