Recertification; How Often To Recertify; Where To Send A Kit For Recertification - Keysight Technologies 85052C User's And Service Manual

3.5 mm precision calibration kit
Table of Contents

Advertisement

Recertification

The following will be provided with a recertified kit:
• a new calibration sticker affixed to the case
• a certificate of calibration
• a calibration report for each device in the kit listing measured values, specifications, and uncertainties
NOTE
A list of NIST traceable numbers may be purchased upon request to be included in the
calibration report.
Keysight Technologies offers a Standard calibration for the recertification of the kit. For more information,
contact the nearest Keysight Technologies sales or service office. See

How Often to Recertify

The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification
depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a
different recertification interval that reflects the usage and wear of the kit.
NOTE
The recertification interval should begin on the date the kit is first used after the
recertification date.

Where to Send a Kit for Recertification

Contact Keysight Technologies for information on where to send your kit for recertification. See
Keysight" on page
5-5. Refer to
your kit.
85052C
"Returning a Kit or Device to Keysight" on page 5-4
Performance Verification
Recertification