Keysight Technologies 85059V Operating And Service Manual
Keysight Technologies 85059V Operating And Service Manual

Keysight Technologies 85059V Operating And Service Manual

1.0 mm precision verification kit
Table of Contents

Advertisement

Quick Links

Keysight Technologies
85059V
1.0 mm Precision
Verification Kit
Operating and
Service Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 85059V and is the answer not in the manual?

Questions and answers

Summary of Contents for Keysight Technologies 85059V

  • Page 1 Operating and Service Manual Keysight Technologies 85059V 1.0 mm Precision Verification Kit...
  • Page 2 THESE TERMS, THE WARRANTY Keysight shall be under no obligation TERMS IN THE SEPARATE to update, revise or otherwise modify © Keysight Technologies, Inc. the Software. With respect to any AGREEMENT WILL CONTROL. 2003-2017 technical data as defined by FAR Technology Licenses 2.101, pursuant to FAR 12.211 and...
  • Page 3: Table Of Contents

    Contents 1. General Information Chapter One at-a-Glance........... . 1-1 About this Manual .
  • Page 4 Returning a Kit or Device to Keysight Technologies ....... .
  • Page 5: General Information

    85059V Operation and Service Guide General Information Chapter One at-a-Glance About this Manual This manual describes the 85059V verification kit and provides replacement part numbers, specifications, and procedures for using, maintaining and troubleshooting this kit. Chapter One at-a-Glance Section Title Summary of Content “Verification Kit Overview”...
  • Page 6: Verification Kit Overview

    5-5 Calibration Definitions The 85059V verification kit is intended to be used with the 85059V 1.0 mm calibration kits. Prior to performing a calibration with your PNA, the calibration kit must be selected and the calibration definitions for the devices in the kit installed in the PNA.
  • Page 7: Installation Of The Calibration Definitions

    Equipment Required but Not Supplied Some items are required or recommended for successful operation of your kit, but are not supplied with the kit. Refer to Table 6-2 for a list of these items and for ordering information. Keysight 85059V Operation and Service Guide...
  • Page 8: Recording The Device Serial Numbers

    1-1. Recording the serial numbers will prevent confusing the devices in this kit with similar devices from other kits. Figure 1-1 Serial Number Engraved on the Nut of a Beaded Line Table 1-1 Serial Number Record for the 85059V Device Serial Number Verification kit ___________________________...
  • Page 9: Clarifying The Terminology Of A Connector Interface

    Improper connections, resulting from pin depth values being out of specification (see Table 2-2 on page 2-3), or from bad connection techniques, can also damage these devices. Keysight 85059V Operation and Service Guide...
  • Page 10: Regulatory And Environmental Information

    (Reference WEEE Directive, 2002/96/EC). China RoHS regulations include requirements related to packaging, and require compliance to China standard GB18455-2001. This symbol indicates compliance with China RoHS regulations for paper/fiberboard packaging. Keysight 85059V Operation and Service Guide...
  • Page 11: Specifications

    Keysight 85059V 1.0 mm Precision Verification Kit Operation and Service Guide Specifications Information in This Chapter Chapter Two at-a-Glance Section Title Summary of Content “Environmental Requirements” on Describes the operating and storage environmental conditions for your page 2-2 calibration kit.
  • Page 12: Environmental Requirements

    For example, if the calibration is performed at +20° C, the error-corrected temperature range is +19° to +21° C. It is then appropriate to perform measurements and performance verifications even though +19° is outside the operating temperature range of the calibration devices. Keysight 85059V Operation and Service Guide...
  • Page 13: Mechanical Characteristics

    — Protrusion is the condition in which the center conductor extends beyond the outer conductor mating plane. — Recession is the condition in which the center conductor is set back from the outer conductor mating plane. Keysight 85059V Operation and Service Guide...
  • Page 14: Center Conductor Protrusion And Pin Depth

    They are, however, important supplemental characteristics related to electrical performance. Keysight Technologies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any center conductor protrusion and have proper pin depth when the kit leaves the factory.
  • Page 15: Electrical Characteristics

    < 0.70 dB Connector Repeatability –35 dB Table 2-4 Mismatch Verification Standard Typical Electrical Characteristics Characteristic Typical Value @ 0 - 120 GHz Return Loss 5 dB Insertion Loss < 3.0 dB Connector Repeatability –35 dB Keysight 85059V Operation and Service Guide...
  • Page 16: Electrical Specifications

    National Institute of Standards and Technology (NIST) through mechanical and electrical paths (for more information on traceability, contact Keysight Technologies. Refer to page 5-5 The factory-measured data for each device is supplied in print and on USB drive with your kit. Keysight 85059V Operation and Service Guide...
  • Page 17: Use, Maintenance, And Care Of The Devices

    Keysight 1.0 mm Precision Verification Kit 85059V Operation and Service Guide Use, Maintenance, and Care of the Devices Information in this Chapter Chapter Three at-a-Glance Section Title Summary of Content “Electrostatic Discharge” on How to protect your instruments and devices against electrostatic page 3-2 discharge.
  • Page 18: Electrostatic Discharge

    4. Remove the short from the cable. Figure 3-1 shows a typical ESD protection setup using a grounded mat an wrist strap. Refer to Table 6-2 for information on ordering supplies for ESD protection. Keysight 85059V Operation and Service Guide...
  • Page 19 Use, Maintenance, and Care of the Devices Electrostatic Discharge Figure 3-1 ESD Protection Setup Keysight 85059V Operation and Service Guide...
  • Page 20: Visual Inspection

    Obvious Defects or Damage Examine the connectors first for obvious defects or damage: — Plating — Bare metal showing — Burrs or blisters — Deformed threads Keysight 85059V Operation and Service Guide...
  • Page 21: Connector Contacts

    Notice the location of the cross hairs in relationship to the center of the figures. Figure 3-2 Contact Integrity Concentricity Figure 3-3 Figure 3-4 show the concentricity of both the male and female 1.0 mm connectors. Inspect the connectors with a minimum magnification of 10X. Keysight 85059V Operation and Service Guide...
  • Page 22: Inspect The Mating Plane Surfaces

    Clean and inspect the connector again if it shows: — Deep scratches or dents — Particles clinging to the mating plane surfaces — Uneven wear Keysight 85059V Operation and Service Guide...
  • Page 23: Inspect Female Connectors

    Inspection of connectors is particularly important when mating nonprecision to precision devices. Supplies and Equipment Needed The supplies and equipment needed to perform the cleaning procedure, and their Keysight Technologies part numbers are listed in Table 7-1, “ Replaceable Parts”, on page 2 and page 7-3...
  • Page 24: Cleaning Connectors

    Cleaning Connectors Supplies and Equipment Needed The supplies and equipment that are needed to perform the cleaning procedures, and their Keysight Technologies part numbers are listed in Table 7-1 on page 7-2 and page 7-3. Basic Cleaning Procedure 1. Use Compressed Air or Nitrogen Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.
  • Page 25 Always completely dry a connector before you reassemble or use it. 4. Reinspect Inspect the connector again to make sure that no particles or residue are present. Keysight 85059V Operation and Service Guide...
  • Page 26 Use, Maintenance, and Care of the Devices Cleaning Connectors Figure 3-5 Cleaning Illustration 3-10 Keysight 85059V Operation and Service Guide...
  • Page 27: Making A Connection

    Good connections require a skilled operator. Instrument sensitivity and coaxial connector mechanical tolerances are such that slight errors in operator technique can have a significant effect on measurements and measurement uncertainties. The most common cause of measurement error is poor connections. Keysight 85059V Operation and Service Guide 3-11...
  • Page 28: How To Make A Connection

    Use very light finger pressure (no more than 2 inch–pounds of torque). 7. Relieve any side pressure on the connection from long or heavy devices, or cables. This assures consistent torque (refer to “Using the Torque Wrench” on page 3-14). 3-12 Keysight 85059V Operation and Service Guide...
  • Page 29: Connector Misalignment

    Turn the connector nut, not the device body. Major damage to the center conductor can occur if the device body is twisted. 1. Use an open-end wrench to prevent the device body from turning. Keysight 85059V Operation and Service Guide 3-13...
  • Page 30: Using A Torque Wrench

    (180 degrees apart) will cause a lifting action (see Figure 3-9 on page 3-15). This lifting action can misalign, and stress the connections of the devices involved. This is especially true when several devices are connected together. 3-14 Keysight 85059V Operation and Service Guide...
  • Page 31 3. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-10 on page 3-16. Wide separation of the wrenches produces a larger residual lateral force on the structure of connected devices. This can degrade connector repeatability. Keysight 85059V Operation and Service Guide 3-15...
  • Page 32 Using the Torque Wrench 5. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its internal pivot point. See Figure 3-11. Do not tighten the connection further. 3-16 Keysight 85059V Operation and Service Guide...
  • Page 33 Do not twist the head of the wrench relative to the outer conductor mating plane. If you do, you apply more than the recommended torque. Keysight 85059V Operation and Service Guide 3-17...
  • Page 34: Handling And Storage

    — Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. 3-18 Keysight 85059V Operation and Service Guide...
  • Page 35: Performance Verification

    Introduction The performance of your verification kit can only be verified by returning the kit to Keysight Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.
  • Page 36: How Keysight Verifies The Devices In Your Kit

    These two steps establish a traceable link to NIST for Keysight to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Keysight Technologies. Recertification The following will be provided with a recertified kit: —...
  • Page 37: Where To Send A Kit For Recertification

    Performance Verification Where to Send a Kit for Recertification Where to Send a Kit for Recertification Contact Keysight Technologies for information on where to send your kit for recertification. Contact information is listed on . Refer to “Returning a page 5-5 Kit or Device to Keysight Technologies”...
  • Page 38 Performance Verification Where to Send a Kit for Recertification Keysight 85059V Operation and Service Guide...
  • Page 39: Troubleshooting

    Keysight 1.0 mm Precision Verification Kit 85059V Operation and Service Guide Troubleshooting Information in This Chapter Chapter Five at-a-Glance Section Title Summary of Content “Troubleshooting Process” on How to troubleshoot a calibration problem. Includes a flow page 5-1 troubleshooting diagram.
  • Page 40 Troubleshooting Troubleshooting Process Figure 5-1 Troubleshooting Flowchart Keysight 85059V Operation and Service Guide...
  • Page 41: Compatible Network Analyzers

    3. Under the heading of your network analyzer, click on the Support tab and then the Document Library icon. 4. Click on the document PDF you want to view. If you need additional information, see “Contacting Keysight” on page 5-5. Keysight 85059V Operation and Service Guide...
  • Page 42: Returning A Kit Or Device To Keysight Technologies

    Returning a Kit or Device to Keysight Technologies Returning a Kit or Device to Keysight Technologies If your kit or device requires service, contact Keysight Technologies for information on where to send it and provide the following: — your company name and address —...
  • Page 43: Contacting Keysight

    Keysight web site: — Go to www.keysight.com. — Enter the document’s part number (located on the title page) in the Search box. — Click Search — Open the PDF and print the document. Keysight 85059V Operation and Service Guide...
  • Page 44 Troubleshooting Contacting Keysight Keysight 85059V Operation and Service Guide...
  • Page 45: Replaceable Parts

    4-2. IMPORTANT! Keysight’s data retention policy is to hold verification data for 18 months. If it has been >18 months since your last recertification, your 85059V kit must be recertified. If you lose this data, refer to https://service.keysight.com/infoline/public/default.aspx or contact Keysight.
  • Page 46: Replaceable Parts

    Replaceable Parts Replaceable Parts Replaceable Parts Table 6-1 lists the replacement part numbers for items included in the 85059V verification kit and Figure 6-1 illustrates the attenuators and beaded lines. Table 6-2 lists the replacement part numbers for items not included in the verification kit that are either required or recommended for successful operation of the kit.
  • Page 47 Items Not Included in the Verification Kit 2 x 4 ft conductive table mat and 15 ft ground wire 9300-0797 ESD heel strap (for conductive floors) 9300-1308 Connector Cleaning Supplies Isopropyl alcohol Cleaning swabs 9301-1243 Keysight 85059V Operation and Service Guide...
  • Page 48 Replaceable Parts Replaceable Parts Keysight 85059V Operation and Service Guide...
  • Page 49 See calibration definitions integrity definitions constants, calibration, See calibration definitions entering contacting Keysight Technologies permanently stored cord kits, intended to be used grounding, part number report standard temperature damage case...
  • Page 50 Keysight Technologies definition of contacting protrusion recession 85059V Index...
  • Page 51 See ESD wrist strap storage 3-18 part number storage case, part number strap heel and wrist, part number supplemental characteristics tag, service temperature affect on electrical performance environmental requirements specifications what to watch out for terminology, connector 85059V Index...
  • Page 52 Index 85059V Index...
  • Page 53 Installation Note Xxxxx-xxxxx...
  • Page 54 This information is subject to change without notice. © Keysight Technologies 2016-2017 Edition 1, June 2017 85059-90006 www.keysight.com...

Table of Contents