Press ENABLE/DISABLE to turn the auto restart feature on/off (Manual mode, network mode and
memodose feature only).
Do not use auto restart for more than 12 times in 24 hours. We recommend
remote control where a high number of starts is required.
If the Auto restart feature is enabled it may cause the pump to start as soon
as power is turned on.
Auto-restart only affects operation of manual mode, network mode and
memodose mode.
If Auto restart is enabled the "!" Symbol is shown on the screen to warn users
that the pump may operate without any manual intervention (pump resumes
with previous settings).
Do not use auto restart for more than 12 times in 24 hours. We recommend
remote control where a high number of starts is required.
If the pump is configured to network mode or analog mode it will respond to
remote commands at any time including immediately after power on. The "!"
Symbol is shown on the screen to warn users that the pump may operate
without any manual intervention (e.g. a remote setpoint could start the pump
with no key press required).