Adjustments and Performance Verification

Calibration Software and specified test equipment is required to perform the adjustments, and can
be used to automate the performance verification testing.
Information on how to obtain this software can be found at
http://www.keysight.com/find/calibrationsoftware
1. Perform the following adjustments, located in the TME field software.
— Absolute Amplitude (Option BBA) Adjustment
2. Perform the following performance verification tests using the TME field software:
— Absolute Amplitude Accuracy at 250 kHz (Option BBA)
— Frequency Response Below 250 kHz (Option BBA)
— Frequency Response Above 250 kHz (Option BBA)
— Channel Match (Option BBA)
A full calibration is required to assure the instrument meets all specifications
IMPORTANT
The end user must ultimately determine whether they want a full calibration to be performed after
the installation of this upgrade or not. If a full calibration is required, arrangements regarding the
level of calibration must be made between the end user and the calibration provider.
If you require further assistance, please contact the Keysight support team.
Online assistance:
http://www.keysight.com/find/assist
If you do not have access to the Internet, contact your local Keysight Technologies Sales and
Service Office, or if in the United States, call 1-800-829-4444
(8 am - 8 pm ET, Monday - Friday).
Installation Note N9020-90257
Option BBA, Analog Baseband I/Q Inputs Retrofit Kit
29

Advertisement

Table of Contents
loading

Table of Contents