1.2. Specifications

Output:
Accuracy:
Voltage Capability:
Regulation:
Output polarity:
Out of limit warning:
Output stability:
Output noise:
Null sensitivity:
Power supply:
Dimensions:
Weight:
1024 Technical Manual
4
0-100mA in 5 ranges
0-99.999mA in 1uA steps
0-9.9999mA in 100uA steps
0-999.99uA in 10nA steps
0-99.999uA in 1nA steps
0-9.9999uA in 0.1nA steps
+/- 0.02% of setting, +/- 0.005% of range, +/- 0.2nA
15V with new batteries or mains power
(11V with minimum allowable battery level)
Load: Better than 5ppm per Volt
Supply: Better than 5ppm per Volt
Positive or negative switch selected. A centre off position
provides an open circuit on the output terminals
A front panel indicator provides a warning of insufficient
drive voltage
Less than 30 ppm per °C (0°C to + 50°C)
Less than 10 ppm per hour at constant temperature
Less than 75 ppm per 6 months
100, 10 and 1mA ranges: less than 5ppm of full scale
100 and 10uA ranges: less than 10 ppm of full scale +/- 0.1nA
Adjustable from +/- 25 mA to +/- 25uA FSD via front panel
control. Maximum resolution is 0.5uA.
Time Electronics power unit type PU2 which is housed in the
rear of the 1024. The PU2 will power the 1024 direct from the
mains or an internal rechargeable battery. The battery is
automatically charged when mains power is connected.
L 220mm x H 160mm x D 200mm
3.3 Kg