2. Pre-Installation

Instrument Description

The GX-274 ASPEC™ is a single-probe X/Y/Z liquid handler that when paired with a VERITY® 4060 Single Syringe
Pump or VERITY® 4260 Dual Syringe Pump can automate liquid handling procedures and solid phase extraction (SPE)
procedures that use positive pressure elution for cartridge-based applications.
For technical specifications for the GX-274 ASPEC, refer to the Chapter 1, Introduction in the GX-274 ASPEC™ User's
Guide.
Unpacking
When unpacking the instrument, check the contents against the information listed under the Unpacking heading in
Chapter 1, Introduction in the GX-274 ASPEC™ User's Guide.
Retain all packing material so the GX-274 ASPEC may be shipped safely in the future.

Installation Site Requirements

Technical Specification
Power Requirements
Temperature Range
Dimensions
(W x D x H)
GX-274 ASPEC™ Installation Qualification
Definition
50 to 60 Hz
Voltage: 100 to 240V (Universal Input)
2A for 100 to 120V
Current
Rating
1A for 220 to 240V
Power Consumption: 250W (Maximum)
5C° to 40°C
59.7 x 54.1 x 57.1 cm (23.5 x 21.3 x 22.5 in.)
Dimensions do not include the VERITY® 4260 Dual Syringe
NOTE
Pumps.
The height dimension does not include the Z-arm, which
NOTE
will vary depending on where the Z-arm is clamped.