ELECTRODE MENU

ELECTRODE STABILIZE

Electrode Stabilization is done when the electrodes are replaced with a new set of EI-23
electrodes or if the K33A is actively used in extreme altitude conditions and a degradation
of splice quality is occurs. If the electrodes need to be stabilized initiate the process as
stated below.
Place SM fi ber in the K33A using fi ber holders or fi ber clamps. Select "Electrode Stabilize".
78 | SWIFT K33A ALL IN ONE FUSION SPLICER
Click
The K33A will perform 60 arc cycles to calibrate the electrodes and the display will count
them down from 60 to zero.
Upon successful completion the message below will be displayed on the touch screen.
NOTE:
Upon successful completion to the Electrode Stabilization a5rn ARC Calibration should
immediately be performed to complete the process see Section 6.6 for instructions.
americanilsintech.com | 79