SET VALUE
DESCRIPTION
A short cleaning discharge is conducted to
Cleaning Discharge
remove fine dust on optical fiber's surface
Amount
upon initial stage of optical fiber arrayal. It
sets intensity of cleaning discharge.
Cleaning Discahrge
It sets time for cleaning discharge.
Clearance
Upon final alignment it examines the gap
sets between both optical fibers.
It sets location of optical fiber spliced at
the center of discharge.
When MFD of both optical fibers differ the
Clearance Location
fusion should apply more arc to the smaller
MFD optical fiber To heat the smaller MFD
optical fiber more, splice loss can be
lowered by moving the clearance location
toward the bigger MFD optical fiber at the
center of discharge.
Sets the initial flash burn discharge level to
clean the fibers prior to the final optical
alignment. If the value of initial discharge
Initial Discharge
amount is too low the cleave angle of the
Amount
fiber is poor an offset error can occur. If the
discharge level is too high the fiber ends
can be burned too much or made round
and Consequently the splice loss value can
be high.
It sets initial time from the beginning of
discharge prior to the final alignment. The
Initial Discharge Time
initial discharge time relates to the amount
of time the flash burn to clean the fiber will
discharge.
56 | SWIFT K33A ALL IN ONE FUSION SPLICER
SET VALUE
GENERAL
USER
MODE
MODE
Editable
Overlap
Discharge Amount 1
Discharge Time 1
Discharge Amount 2
Editable
Editable
Discharge Time 2
Time Period for
Discharge 2 Being On
Time Period for
Discharge 2 Being Off
DESCRIPTION
It sets duplication of optical fiber on optical
fiber's advance amount. If [initial discharge
amount] is weak or [initial discharge time] is
short, set the [overlap] somewhat small and
if discharge amount is strong and the time
is long, set it somewhat big.
Main discharge can be adjusted in 2 levels.
The first level of discharge is [discharge
amount 1] and the second is [discharge
amount 2]. [Discharge amount] 1 is set in
this area.
It sets time for [discharge amount 1].
[Discharge amount 2] is the second level of
discharge. [Discharge amount 2] is set in
this area.
It sets time for [discharge amount 2].
It sets time for [discharge amount 2].
[Discharge time 2] is generally set as "OFF."
It can set the discharge time as a very long
time period but when [discharge time 1]
and [discharge time 2] exceed 30 seconds,
electrode can be damaged.
While [discharge amount 2] is on discharge,
you can set the discharge amount as ON
and OFF in turn. Time period for [discharge
amount 2] being on is set in this area. For
re-discharge, set the discharge time as
"ON" at all times.
It sets time period for discharge of
[discharge amount 2] is turned off. When
[discharge amount 2] is occasionally
stopped, re-discharge can also be stopped.
When re-discharge is continuously
required, set as "OFF."
GENERAL
USER
MODE
MODE
Non-
Editable
Editable
americanilsintech.com | 57