This product is not made with natural rubber latex
Intended Use:
The single use Quick Catch™ in-line suction polyp trap is used for suction retrieval of endoscopically removed polyps.
Warnings and Precautions:
Consult the medical literature relative to techniques, complications and hazards prior to the performance of any endoscopic
procedure.
The Quick Catch™ in-line suction polyp trap has a built-in luer lock that fits standard luer lock syringes (See Figure 2). This allows
for the use of irrigation to remove specimens within the specimen screen.
When removing a specimen from the trap, make sure that the entire specimen is located within the specimen screen prior to
removing trap from system to prevent the loss of retrieved tissue.
Endoscopic procedures should only be performed by persons having adequate training and familiarity with endoscopic techniques
and polyp retrieval.
Do not attempt to reuse, reprocess, refurbish, remanufacture or resterilize this device. STERIS Endoscopy did not design this
device, nor is it intended to be reused, reprocessed, refurbished, remanufactured, or resterilized. Performing such activities on
this disposable medical device presents a safety risk to patients (i.e. compromised device integrity, cross-contamination,
infection).
Contraindications:
Contraindications include, but are not limited to, those specific to any endoscopic procedure.
Prior to Use:
1.
Prior to clinical use, inspect and familiarize yourself with the device. If there is evidence of damage (i.e. deformed, cracked, kinked
tubing, damaged packaging), do not use this product and contact your local Product Specialist.
Directions for Use (Specimen Retrieval):
1.
Prior to clinical use, familiarize yourself with the device and read all the Instructions for Use.
2.
Connect the luer fitting of the Quick Catch™ in-line suction polyp trap to the suction system (See Figure 1).
3.
Connect the tube on the Quick Catch™ in-line suction polyp trap to the suction connector (barb) on the endoscope (See Figure 1).
4.
Retrieve resected polyp by applying suction via the endoscope suction valve.
5.
Remove the Quick Catch™ in-line suction polyp trap from the endoscope and the suction system tubing once the polyp is retrieved.
6.
Reattach the suction system tubing to the suction connector (barb) on the endoscope.
7.
Separate the trap by rotating both halves in opposite (counterclockwise) directions (See Figure 2).
8.
Retrieve the polyp by removing it from the specimen screen and placing it in the appropriate fixative for histopathology in accordance
with institutional guidelines.
Product Disposal:
After use, this product may be a potential biohazard which presents a risk of cross-contamination. Handle and dispose of in
accordance with accepted medical practice and applicable local, state and federal laws and regulations.
Issued Date: October 2019
Warning:
An issued or revision date for these instructions is included for the user's information. In the event two years has elapsed between this date and product
use, the user should contact STERIS to determine if additional information is available.
Unless otherwise indicated, all marks denoted with ® or ™ are registered with the U.S. Patent and Trademark Office, or are trademarks
owned by STERIS Corporation.
Serious incidents that have occurred in relation to this medical device should be reported to the manufacturer and competent authority in the
country where the incident occurred.
US Endoscopy, a wholly owned subsidiary of STERIS Corporation.
Made in the U.S.A.
00732560 Rev E