Settings; Calibration Menu - OHAUS Valor 1000 Series Instruction Manual

Hide thumbs Also See for Valor 1000 Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
TM
Valor
1000 Series

4. SETTINGS

The scale allows the user to customize several settings including calibration,
backlight operation, unit change and auto shut-off. Long press Menu for more than
3 seconds until you see C.A.L. Release the Menu key after you enter the user menu.

4.1 Calibration Menu

Enter this menu to perform calibrations.
Once the scale displays
perform either a Span or Linearity calibration.
For legal-for-trade use, this may have to be done in accordance with local weights
and measures requirements. Contact your local servicing scale dealer for more
information.
Press Tare to exit Calibration and to move on to the Setup Menu.
Note: Calibration is not permitted in some regions or when the scale is in Approved
mode.
SPAN
SPAN [
]
Initiates a span calibration procedure using a span calibration weight equal to the
full capacity of the scale. Press On/Off to enter the SPAN calibration routine, or
press Tare to proceed to a Lin calibration.
When 0.000kg is shown, short press On/Off to capture the zero weight if the
platform is empty.
When the full capacity of the scale is shown, place a weight equal to the capacity of
the scale and short press On/Off.
If the scale displays "
scale did not complete calibration.
To exit calibration at any point, short press Tare/Menu.
LIN
Lin [
]
Initiates a linearity calibration procedure (zero, mid-point and span).
When performing a linearity calibration, the scale will prompt you for a zero weight -
- the platform without anything on it and weight display at zero -- a Midpoint weight
equal to 50% of the Span weight, and a Span weight equal to the full capacity of the
scale.
When 0.000kg is shown, short press On/Off to capture the zero weight if the
platform is empty.
When the half capacity of the scale is shown, place a weight equal to 50% of the
scale's capacity on the pan and short press On/Off.
When the full capacity of the scale is shown, place a weight equal to the capacity of
the scale and short press On/Off.
If the scale displays "
scale did not complete calibration.
To exit calibration at any point, short press Tare/Menu.
C.A.L
, press On/Off to enter the Calibration Menu and
Cal-E
", one of the weights used to calibrate is wrong and the
Cal-E
", one of the weights used to calibrate is wrong and the
EN-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents