6. MAINTENANCE

6.1 Cleaning

The housing may be cleaned with a cloth dampened with a mild detergent if
necessary.

6.2 Troubleshooting

The following table lists common problems and possible causes and remedies. If
the problem persists, contact OHAUS or your authorized dealer.
Symptom
Possible Cause
Calibration error: unstable
CAL E
environment or incorrect calibration
weight
The Power On weight on the platform
Err 8.1
exceeds the zero range.
The Power On weight on the platform
Err 8.2
is below the zero range.
Over load (weight exceeds rated
Err 8.3
capacity)
Err 8.4
Under load (pan removed)
Err 8.5
Tare weight out of range
Internal data error or calibration data
Err 9.5
error
Err 13
Failure to write to EEPROM
Lo.Bat
Battery very low
--NO--
Action not allowed
------
Busy
WARNING: Electric Shock Hazard. Disconnect the equipment from
the power supply before cleaning.
Make sure that no liquid enters the interior of the instrument.
Attention: Do not use solvents, harsh chemicals, ammonia or
abrasive cleaning agents.
TABLE 5-1. TROUBLESHOOTING.
TM
Valor
1000 Series
Possible Solution
Check environment and
calibration weights, and
then re-calibrate.
Clear pan.
Check if pan is missing.
Remove weight from pan.
Re-install pan.
Tare value exceeds
maximum.
Contact OHAUS or your
authorized dealer for help.
Contact OHAUS or your
authorized dealer for help.
Replace batteries; Charge
rechargeable battery.
/
/