Troubleshooting
Troubleshooting Using Diagnostic Test
Program Overall
Instrument
Information
Test Conditions
Test Items
Figure 1-15 Test dialog box
Keysight E5072A Network Analyzer
The option and serial number for the E5072A, and test date
can be entered automatically.
The test conditions, "Temperature", "Humidity" and person
doing the test ("Tested by") can be entered.
The following test can be selected. To execute the test you
select, click "Execute" button.
•
DC-BUS Test
•
Internal Level Monitor Test
•
RF Output Level Range Test
•
RF Output Level Power SweepTest
•
Receiver Absolute Measurement Test
•
Receiver Compression Test
•
Receiver IF Ranging Test
When each test is executed, the test dialog box is displayed.
The following screen is an example of the test dialog box.
To execute the test, click "Execute" button. To return the
E5072A Diagnostic test program dialog box (Main Menu),
click "Back to Main Menu".
The test program will prompt you to connect cables or Short
termination to the E5072A. Follow the instructions as shown
below on the E5072A display for performing the test.
25