Honeywell XNX-AM Series Technical Manual page 120

Hide thumbs Also See for XNX-AM Series:
Table of Contents

Advertisement

XNX Universal Transmitter
EC Sensor Cartridge and Cell Replacement
CAUTION
!
Caution: For toxic ells, remove the shorting clip from the bottom of the cell prior to installation.
No shorting clip is provided with oxygen cells..
The serviceable sensor allows replacement of the cell inside the
sensor. The cell must be replaced with one of the same type as
the original.
When replacing oxygen (O
is between 10 and 15 minutes. This warm-up is required only
after sensor cell replacement.
Replacing the Sensor Cell
To replace the cell follow this procedure:
1. Unscrew the weatherproof cover, loosen the sensor
retainer locking screw, and unscrew the sensor retainer.
2. Remove the old sensor by pulling without twisting.
3. Unscrew the sensor cap.
4. Remove the old cell by pulling without twisting.
5. Verify that the new cell is the same type as the old one.
6. Plug the new cell into the sensor. Align the sensor pins
with the connector holes in the PCB.
7. Refit the sensor into the transmitter ensuring that the pins
are fully aligned before gently pushing it fully home.
8. Sensor warm-up will begin and the display will alternate
between two screens: "Fault 151" and "WARM."
Maintenance
) sensor cells, the initial warm-up time
2
Figure 226. Sensor screens during warmup
9. Follow the procedure to accept the new sensor in the
Accept New Sensor Type
10. Recalibrate the sensor following the procedures in the
Span Calibration for EC/mV Sensors, and Searchpoint Optima
section.
Transmitter
Old Sensor
3
Sensor Retainer &
Locking Screw
2
1
EC Sensor Removal
Figure 227. XNX EC sensor and cell replacement
118
section.
New cell
Sensor Cap
New Sensor
Sensor Cell Replacement
4
5
6
Zero and