Setup Guide
Option C: copying coefficient values from another sensor
The sensor to which you are copying the values:
•
Must have plate reference done
•
Must be used in the same area as the sensor from which you copy the values
Recommended to copy dry road coefficient values from one sensor to another when dry road
reference has been done on a similar road surface before and surface coefficients are known.
1.
Check the values from a sensor that you are using in the same area. Use the following message:
GET PARAMETER
The requested parameters are 0x53, 0x54, and 0x55.
2.
For the sensor that you want to calibrate, give the values for the 3 parameters (Road coefficient,
laser 1, laser 2, and laser 3). Use the following message:
SET ROAD COEFFICIENTS
M212169EN-D
23