STEP 5. Verify Operation of the Network Analyzer Modules

STEP 5. Verify Operation of the Network Analyzer Modules

The Operator's Check

Allow the network analyzer modules to warm up for at least 15 minutes before performing
NOTE
the Operator's Check.
The operator check is a self test that checks the basic operation of the
assemblies in all of the measurement port signal paths. This allows you to
confirm the cable connection and modules' basic operations. No tools are
required to execute this test.
Removing the DUT from the test port is recommended.
All the test ports will be short circuited during the test. If the test fails, an error
dialog box is displayed. Click on Help to show the troubleshooting hint in the
help file.
The operator's check is not complete solution to find all malfunctions. Hence,
perform the verification and performance tests to check the complete
performance of M9485A.
Operator's check is executed at the Soft Front Panel boot up default. You can
turn off the start up test by the preference setting (System > System Setting >
Preference...).
Optionally, you can execute the test manually by selecting System > Service >
Operator's Check.