■ External Control Signals (15-pin D-sub type)
Pin No.
Signal name
External
1
+12 V
view
2
Lamp change signal
3
Lamp ON signal
4
Lamp STABLE signal
5
Shutter OPEN signal
6
Output intensity upper limit signal This signal is output when the output intensity reaches the upper limit.
7
Alarm signal
8
Output intensity lower limit signal This signal is output when the output intensity reaches the lower limit.
9
Output intensity UP
10
Lamp OFF trigger
11
Lamp ON trigger
12
Shutter auto trigger
13
Shutter OPEN / CLOSE
14
Output intensity DOWN
15
GND
Note
Signal output w ~ i
Signal input o ~ !4
● To use the external control function, use the plug for the ex-
ternal control signal supplied with the main unit.
Insert the plug into the signal I/O connector on the back of the
main unit, and secure it with the left and right screws.
Plug for external control signal
D-sub type
Model name
■ Signal Output
The circuit between the signal output pin and the GND pin
(see the above table) turns ON to output signals indicating
each condition. The load connected to the output must meet
the following conditions.
• Voltage: Maximum 50 V
• Current: Maximum 30 mA
Internal circuit at
output side
Photocoupler
Signal
open collector
100 Ω
output pin
Caution
● Although a protective resistance (100 Ω) is included
in the output transistor, also put a current limit resis-
tance (1 kΩ min.) in the external circuit so as to pre-
vent a current exceeding the rated level from flowing.

External Control

+12V dc output (100mA max.)
This signal is output when the accumulated lamp operation time reaches 4000 hours
(L8253:3000 hours).
This signal is output when the lamp ON.
This signal is output about 5 minutes after the lamp ON.
This signal is output while the shutter open.
This signal is output when an alarm indicating overheat, lamp error, shutter error or
lamp change is activated.
When this signal is inputted while the shutter is open using the manual mode, the
output intensity increases (the diaphragm opens) as long as it is inputted.
Momentary (about 0.5 sec.) input of this signal makes the lamp turn off.
Momentary (about 0.5 sec.) input of this signal makes the lamp turn on.
Momentary (about 0.5 sec.) input of this signal makes the shutter operate in the same
way as when the shutter button in auto mode is pressed once. (The shutter opens/
closes as set in the program.)
When this signal is inputted, the shutter opens as long as it is inputted. When no signal
is inputted, the shutter closes.
When this signal is inputted while the shutter is open using the manual mode, the
output intensity reduces as long as it is inputted. (The diaphragm closes.)
COM of signal output/signal input
Open collector output using a photo coupler. Each terminal to GND turns on and out-
puts each condition.
A signal is inputted with the signal input pin and GND turned on (shorted). When indi-
cated "trigger", momentary (about 0.5 sec.) ON is provided.
15P connector
XM3A-1521 plug (manufactured by Omron) compatible parts
XM2S-1511 hood (manufactured by Omron) compatible parts
External circuit
50 Vmax.
1 kΩ min.
GND pin
Description
signal I/O connector
■ Signal Input
Each operation is possible by turning ON the circuit between
the signal input pin and the GND pin (see the above table).
The contact or transistor used to implement the short must
be rated as follows.
• Withstand voltage: 30 V minimum
• Current: 50 mA minimum
Internal circuit at
input side
+12 V
1 kΩ
Photocoupler
ON current =
approx. 10 mA
23
screw (M2.6)
screw (M2.6)
External circuit
Signal
input pin
GND pin
Contact or photocoupler