Specifications; Thermocouple Mv Input; Thermocouple Mv Output; Thermocouple Ma Input - Fluke 714C User Manual

Thermocouple calibrator
Hide thumbs Also See for 714C:
Table of Contents

Advertisement

714C
Users Manual

Specifications

Specifications are based on a one year calibration cycle
and apply from +18 °C to +28 °C unless stated otherwise.
All specifications assume a 5 minute warmup period.

Thermocouple mV Input

Range
-10 mV to 75 mV
Temperature coefficient: ±(0.002% of reading + 0.002 % of range) /°C ( <18°C or >28°C)

Thermocouple mV Output

Range
-10 mV to 75 mV
Temperature Coefficient: ± (0.002 % of output + 0.002 % of range) /°C ( <18 °C or >28 °C)
mA Input
Range
0 mA to 24 mA
Temperature Coefficient: ± (0.002 % of output + 0.002 % of range) /°C ( <18 °C or >28 °C)
28
Resolution
0.001 mV
Resolution
0.01 mV
Resolution
0.001 mA
Accuracy (% of Reading + Floor)
1 year
0.015 % + 10 µV
Accuracy (% of Output +Floor)
1 year
0.015 % + 10 µV
Accuracy (% of Reading + Floor)
1 year
0.01 % + 2 µA
2 year
0.02% + 15 µV
2 year
0.02 % + 15 µV
2 year
0.02 % + 4 µA

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents