Symptom/Error Message
No last response code given
No status lights when SoloPORTER is first
plugged in
Red light blinks whilst green light is solid
Red light blinks after green light goes off
Solid red for 5 seconds, followed by solid green
and flashing red for 8 seconds
Copyright © 2008-2012 Sentek Pty Ltd All rights reserved
G
G
O
O
T
T
I
I
N
N
G
G
U
U
I
I
D
D
E
E
Possible Cause of Failure
Communication problem between SoloPORTER
and probe
Both SoloPORTER and Solo Head Unit batteries
are too low
Low Solo Head Unit batteries
Low SoloPORTER battery
No USB flash drive in SoloPORTER
USB flash drive is write protected
Wires disconnected from probe
Wires connected to wrong position
Probe not configured correctly
A battery cap taken off (no Head Unit power)
Incorrect firmware on probe or SoloPORTER
Appendix C - Sentek SOLO Quick Reference Guide
SoloPORTER Troubleshooting Guide
Possible Solution
Check probe cabling, interface and all connections
Replace SoloPORTER and Solo Head Unit batteries
Replace Solo Head Unit batteries
Replace SoloPORTER battery
Insert USB flash drive and confirm is fully inserted into
socket
Unprotect the flash drive or replace with unprotected flash
drive
Make sure wires are configured correctly and not loose
Make sure wires are configured correctly and not loose
Check probe configuration
Either a battery cap taken off or not sealing correctly
Probe requires firmware version 1.2.2 (or later) and
SoloPORTER requires "VDAP0364.FTD" file (or later) in
the SoloDATA directory
Page 22