The Check Configurations are defined below.
NOTE: Default settings are in bold.
CONFIGURE ITEM
OPTION
Check Mode
Check Weighing/Check
Counting
Auto Tare
Off/On/Accept
Accumulate
Off/Manual/Automatic
Audible Signal
Off/Under and
Over/Accept/Under/Over
CHECK COUNTING
1.
Press the
button
and select Check
Mode to Check
Counting.
(See Figure 57)
2. Place objects on pan. The Under/Accept/Over status
is shown in the progress bar area, while the actual
number of pieces is shown on the main display line.
Set the Average Piece Weight (APW)
NOTE: It is recommended that the APW is larger
than 1d. If APW is between 0.05d and 1d, a
warning message will be displayed and the
information line will show "APW is low!". If APW is
less than 0.05d, an error message will appear
and the APW value cannot be stored.
There are three methods to set the APW. See Application
Setup at right.
Defining Over/Under Limits
Press the Edit Limit button
to define counting limit.
(See Figure 58)
wNOTE: See Check Weighing on page 17 for
information on how to set the Over/Under limits.
PAGE 17 OF 33
APPLICATIONS CONTINUED
DESCRIPTION
Check weighing mode
Check counting mode
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a container automatically.
Accept: If object weight is in the range of the Over and Under Limit you set, auto tare will
be performed.
Off: The icon "Σ" does not display.
Manual: The softkey "Σ" displays. User can press the key to accumulate the stable weight.
Automatic: The icon "Σ" displays. The weight will be accumulated automatically.
NOTE: The load to be accumulated has to be >= 10d. The next
accumulation can only start once the pan is cleared.
When LFT is ON (no such limit when LFT is OFF or the approve mod is OIML),
a. Gross weight and net weight cannot be accumulated;
b. Gross 0 has to be reached before a new sample accumulation.
When LFT is OFF,
a. Gross weight and net weight can be accumulated;
b. A new sample can be accumulated after reaching gross 0 or net 0.
Off: No beep.
Under and Over/Accept/Under/Over: Beep when reaching the selected check point.
APPLICATION SETUP
The application can be customized for user preferences
1.
Figure 57
2. Select the list item
3. To return to Application home screen, press the button
Figure 58
Press the softkey
corresponding to the
icon to enter the
Configuration Setup.
Configuration menu is
now displayed.
(See Figure 59)
and press the softkey
corresponding to the
icon, to change
setting. (See Figure 60)
corresponding to the
User
Menu
Figure 59
Figure 60
icon.
1219 IH-8105