Ventilation modes

CPAP mode

CPAP mode delivers a constant level of pressure during inspiration and expiration.
With CPAP administered via a valved breathing system, the inspiratory triggering level is settable to
optimise the control of the exhalation valve and minimise the patient's work of breathing. Adjust trigger
sensitivity to accurately report patient respiratory rate.
In a vented system, the inspiratory triggering level is settable to optimise monitoring and alarm
performance.
Parameter
CPAP (cmH
O)
2
Trigger type
Trigger
Supplementary features:
Apnoea response (Valve circuits only)
CPAP operation with a single limb circuit with intentional leak is displayed.
78
Setting
All circuits:
3.0 to 20.0 [5.0]
Double limb circuit:
Flow/Pressure
Double limb circuit:
When Trigger Type is set to Flow:
Adult: 0.5 to 15 [1.0] (L/min)
Paed: 0.5 to 15 [0.5] (L/min)
When Trigger Type is set to Pressure:
Very Low to Very High [Medium]
Single limb circuit:
Very Low to Very High [Medium]
Single with intentional leak:
Very Low to Very High [Medium]