Verify Secure Placement of Sponge
Gently tug on the sponge. If properly seated in the pedestal, the sponge should
1
not come loose. If it does come loose (Figure 9-19), repeat Steps 1–6 in preceding
section, "Position Sponge in Pedestal.
Other Repairs
Repairs other than electrode replacement in a GSN 200 or sponge replacement in an
HCGSN must be done by a trained EGI Net technician and will require the return of
the Net to EGI for servicing.
In the event that a Net needs servicing or refurbishment, contact EGI Technical
Support (Appendix A). Please disinfect the Net (see page 81) before returning it to
EGI.
Disposal of Electrodes or Sponges
See "Component Recycling and Disposal" on page 56.
Geodesic Sensor Net Technical Manual
Figure 9-19. A poorly seated sponge
S-MAN-200-GSNR-001 • January 31, 2007
9: Replacing GSN Parts
99