chapter 7
R
INSING AND
7
CHAPTER
D
ISINFECTING
T
his chapter describes the processes of rinsing, disinfecting, rinsing again, and
drying the Net, which is performed after each use. Proper handling of the Net
during these stages is important to ensuring the health and longevity of your Net. The
information in this chapter applies to both the GSN 200 and HCGSN models, unless
otherwise noted.
Caution!: Proper care of your Net depends on following the instructions in this
chapter. Failure to do so could result in degradation of your Geodesic Sensor Net.
Rinsing
The following instructions apply to all GSN models: GSN 200s, HCGSNs with sponge
inserts, and HCGSNs without sponge inserts. The HCGSNs without sponge inserts
require some additional care; see "Rinsing (HydroCel HS)" on page 80.
Fill the plastic bucket marked "Electrolyte" about half full with clean, warm tap
1
water. Do not use hot or boiling water. Submerge the sensor end of the Net in
the water.
Gently agitate the Net in the water for 10–20 seconds. Drain the water and refill
2
the rinse bucket. Further agitate the Net, pressing the sponge tips with your
palm to cycle water through them, if applicable. Repeat this rinse-and-drain
process a total of four times.
Remove the Net from the rinse bucket and remove the excess water by patting
3
the sponge tips (or electrode cups, if spongeless) with a clean dry towel.
Geodesic Sensor Net Technical Manual