Cleaning Process

1. Pre-clean with cold water, washing several times.
2. Clean probe with a solution of 5-10% mild detergent for about 10-15 seconds.
3. Continuously clean with cold water to avoid any chemical residues remaining
on the device.
4. Rinse the device with drinking water or distilled water to ensure the removal of
all remnants of cleansing agent.
5. If necessary, repeat step 2 through step 4 in order to make sure the device
is clean.
6. A final wipe down is necessary to avoid water stains.
Disinfecting Process
1. Wipe probe with a cloth dipped in a disinfectant solution such as 70-75%
isopropyl (rubbing) alcohol solution: contact time ≥10 minutes for disinfection.
2. Rinse off disinfectant residue.

Drying Process

• Do NOT drop or crush; this device is not shock proof.
• Do NOT dismantle or modify device other than changing batteries.
• Do NOT store the device in an extreme environment with direct sunlight or
high/low temperatures, or in a location with high humidity or severe dust.