How To Use; Special Features - ADC Adtempt 419 Instructions For Use Manual

Digital hypothermia thermometer
Table of Contents

Advertisement

How to Use

1. Disinfect the probe with isopropyl (rubbing) alcohol before using.
2. Press and release the multifunction button. The display will read
confirming the unit is functioning properly, then will show L°F with °F flashing.
3. We recommend using probe covers with all digital stick thermometers.
If using, place cover over tip of unit.
4. Place the probe in the appropriate position for your measurement site.
(see Ensuring Accurate Temperature Measurement below).
5. Once the °F stops flashing the measurement will be displayed and an alarm
signal will sound.
6. The unit will automatically turn off in about 3 minutes. To prolong battery
life, press the multifunction button to turn off the device manually.

Special Features

Last-Temperature Memory Recall: With the thermometer off, press and hold the
multifunction button for approximately 5 seconds. The display of 188.8 will change to
show the last reading. This reading is displayed as long as the button is held down.
When the button is released, the thermometer will be in "ready-to-measure" mode and
the recalled temperature will be erased.
Switching Temperature Modes: Within 3 seconds of turning on the thermometer, the
Celsius or Fahrenheit mode can be switched by pressing and holding the multifunction
button for at least 2 seconds.
Ensuring Accurate Temperature Measurement
We recommend using probe covers (Figure 1)
with all digital stick thermometers.
Oral Use: Place the probe well under the patient' s tongue
(Figure 2). Keep mouth closed while thermometer is reading.
Rectal Use: Lubricate the probe cover with a water-
soluable jelly for easier insertion; do not use petroleum
jelly, as it may cause an inaccurate reading. Insert the tip
of the probe no more than 1/2 inch into the rectum.
4
Figure 1
Figure 2