Environmental Information

NOTE
Samples of this product have been type-tested in accordance with the Agilent
Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation and end-use; those stresses
include but are not limited to temperature, humidity, shock, vibration,
altitude and power-line conditions. Test methods are aligned with IEC
60068-2 and levels are similar to MIL-PRF-28800F Class 3.
Table 4-1
Environmental Information
Parameter
Temperature
Operating
Storage
Error-corrected range
Altitude
Operating
Relative humidity
Note: In conditions of very high humidity (below 95%) at changing temperature, there is a small
risk of internal condensation that could cause the CPU real-time clock to temporarily stop
incrementing time.
E8356-90001
Safety, Regulatory, and Environmental Information
Required Values/Ranges
0 °C to +40 °C
The instrument powers-up and displays no error
messages within this temperature range (except for
"source unleveled" error message that may occur at
temperatures outside the specified performance
temperature range of 25
−40 °C to +70 °C
23 °C ±3 °C with less than 1 °C deviation from
calibration temperature
0 to 4,600 meters (≈15,000 feet)
Type tested, 0% to 95% at 40 °C, non-condensing
Environmental Information
± 5 °C).
4-7