Table 10-5
Option modules
Optional module 2.2
Table 10-6
Climatic conditions
Permissible oper‐
ating elevation
Permissible ambient temperature
Permissible humidity
Under atmospheric conditions at 0.8 to 1.1 bar
1)
Table 10-7
Cable gland
M20 x 1.5
M20 x 1.5
M20 x 1.5
M25 x 1.5
Restraining capacity in accordance with DIN EN 50262 Class B
IP55 protection is only achieved if the diameters of the inserted cables are within the clamping range
of the cable gland.
Field device
Operating Instructions, 05/2015, A5E35640463-02
Field control unit: Option modules
Field control unit / field module Climatic conditions
Hazardous areas:
1)
Non-hazardous
areas:
Field control unit: Cable gland
10.2 Technical specifications of field device
3 analog outputs 0/4 to 20 mA, isolated; max. load 750 Ω
4 analog inputs 0/4 to 20 mA, non-floating, internal resistance
≤ 100 Ω
2000 m AMSL to EN 60079-0
3000 m above sea level
● -30 to +70 °C during storage and transportation
● +5 °C to +55 °C for regular operation with OXYMAT 7
● +5 °C to +60 °C for operation with OXYMAT 7 and with
limited accuracy
< 90% relative humidity as annual average, during storage
and transportation (must not fall below dew point)
Clamping range diameter
6.1 to 11.5 mm
9.1 to 15.9 mm
12.5 to 20.9 mm
14 to 22 mm
Technical specifications
89