The following applies to the connections:
NC
x/3
x/2
x/1
FM-
XX
Notes on the inputs/outputs:
● Digital outputs (DO): with changeover contacts, freely parameterizable, e.g. for measuring
range identification
Load: 24 V AC/DC/1.7 A, isolated, non-sparking
● Digital inputs (DI): designed for 24 V, isolated, freely parameterizable, e.g. for autoranging
● Analog outputs (AO): 0/4 ... 20 mA, isolated; max. load 750 Ω
● Analog inputs (AI): 0/4 ... 20 mA, non-floating, internal resistance ≤ 100 Ω
Note
Display of connections AO1..3 on the LUI
The connections identified by AO1..3 are displayed on the LUI as AO11..13.
Note
Strain relief of the connecting cables
Make sure that the strain relief of the cable gland is not overloaded. Make sure that the
connecting cables are force-free.
If necessary, install an external strain relief.
Procedure for connecting
1. Open the center cover of the field control unit.
2. Open the cable gland using suitable tools.
Field device
Operating Instructions, 05/2015, A5E35640463-02
AI 2P
AI 3P
AI 4P
NC
NC
CANL
FM-09
NC
GNDPWR
NC
GNDPWR
FM-06
GNDPWR
FM-05
GNDPWR
FM-04
Not connected
Normal closed
Center
Normal open
XX indicates the number present on the core of the connecting cable.
60
AI 2N
61
AI 3N
62
AI 4N
63
NC
64
GND
FM-07
65
CANH
FM-08
66
NC
67
NC
68
P24V
FM-03
69
P24V
FM-02
70
P24V
FM-01
Wiring
6.5 Field control unit