Errors - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis field device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

9.4

Errors

Table 9-1
Error messages
Fault message
Description
Invalid device
No analyzer module was detected. At least one analyzer module is re‐
configuration
quired to use the device.
Firmware ver‐
The firmware of the basic device and the analyzer modules must be com‐
sions incompati‐
patible with one another.
ble
Unknown mod‐
An analyzer module has been detected which is not supported in the
ule type detected
previous firmware version.
Memory error
A hardware fault, faulty firmware or no firmware at all has been detected
detected
in the following analyzer modules, e.g., AM1. Possible causes:
● No firmware is present on the analyzer module.
● The existing firmware is invalid.
● The program memory is faulty.
In this case, the analyzer module starts in bootloader mode and is ready
to perform a firmware update.
If the error still exists despite the firmware update, the analyzer module
PCB must be replaced.
Cable fault de‐
The installed analyzer modules could not be recognized correctly.
tected
Possible causes: Faulty or incorrectly connected CAN bus cable.
Field device
Operating Instructions, 05/2015, A5E35640463-02
Service and maintenance
9.4 Errors
Measures
Install at least one analyzer
module.
Please contact the service de‐
partment.
Please contact the service de‐
partment.
Please contact the service de‐
partment.
Check the CAN bus connec‐
tions.
85

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

Table of Contents