Instructions for Use
CLARUS 500
NOTE
Action
NOTE
2660021165005 Rev. A 2017-08
Events are automatically recorded in 1-5 audit files of 5 Mb each.
When file size limit is reached, the device overwrites existing files.
The default folder for the audit log files is C:\ProgramData\Carl
Zeiss Meditec\...\Logs.
Save (export) audit log files regularly
to ensure events of consequence can be tracked should you en-
counter a data error.
Log on as Administrator.
u
Click Tools > Export Audit Log File.
u
In the Browse folder, browse to the export location.
u
Click Save.
u
ð The log is exported as a .zip file with the format: Audit-
Log_dd_mm_yyyy_hh_mm.
5.3 Preparing the Device
1. Wipe the chinrest and forehead rest with an alcohol pad, and
allow the assembly to dry. (See Cleaning the Chinrest and Fore-
head Support [} 112].)
2. Carefully read and understand any instructions provided by the
officiating physician or researcher.
5.4 Preparing the Patient
Keep the front of the lens clean.
The quality of image capture is greatly affected by the cleanliness
of the front lens.
Avoid touching the lens with anything but approved cleaning
u
equipment.
Use the protective cover to keep dust off the lens when the de-
u
vice is not in use.
For complete instructions on cleaning the lens, refer to: Cleaning
and Disinfection [} 109]
5.4.1 Dilating the Patient's Eyes (Optional)
The minimum pupil size for the CLARUS 500 is 2.5 mm, which is
typically achieved without dilation.
5.4.2 Selecting the Patient
Finding a patient's name can vary depending on the systems you
use.
5 Operation
5.3 Preparing the Device
53 / 144