Description of Questionable Status Bit
Mnemonic
Bit
FE
0
UV
1
UV
2
OVP
3
OCPP
4
OCP
5
OPP
6
OTP
7
LDF
8
Description of Operation Status Bit
Mnemonic
CAL
TRG
Description of Standard Event Status Bit
Mnemonic
Bit
OPC
0
QYE
2
DDE
3
EXE
4
CME
5
Value
Weight
1
2
4
8
16
32
64
128
256
Bit
0
5
Value Bit
Weight
1
4
8
16
32
Copyright © Itech Electronic Co., Ltd.
Bit
Meaning
Below the frequency range or above the
frequency range.
AC voltage is lower than 45V.
DC voltage is lower than 7.5V.
AC voltage is upper than 350V
The peak current value in AC/DC mode is
upper than 18A or 45A.
The current RMS value in AC/DC mode is
upper than 18A.
The power value in AC/DC mode is upper than
1800W or 4500W.
The temperature value is upper than the limit
value.
Load Failure.
Value Bit
Weight
1
32
Meaning
Operation Complete
Query Error
Device Error
Execution Error.
Command parameter is invalid or inconsistent
with electronic load, or the command cannot be
executed.
Command Error. Command syntax is incorrect.