Measurement Setup Parameters
Channels: From the menu Input > Channels, available options are:
1 channel through n channels (n depends on your digitizer model),
I+jQ (IN1+jIN2)
The custom channel configuration can be used to specify a non-standard mapping
between the digitizer input and the VSA logical channel. The default mapping of the
logicals channel to the digitizer inputs is:
Logical Channel
1
2
From the menu Input > Analog the following parameters can be set:
Range: Thedigitizer offers several full scale range. Please refer to
product datasheet for details.
Coupling: DC
Input Impedance: 50 Ohms.
Connection: Single Ended.
The frequency measurement parameters can be configured from the MeasSetup >
Frequency menu.
Frequency: The default setup is shown below as example.
Model Option
-SR2
Span: The maximum span is dependent upon both the sampling rate and
the analog bandwith of the particular digitizer model.
Model Option
-SR2 -F05
-SR2 -F10
The maximum span is obtained with following formula:
Maximum Span = Min( Sample Rate / 2.56 ; Analog Bandwidth)
The analog bandwidth depends on the digitizer, the option (-Fxx) and interleaving
mode (see product datasheet for details).
Maximum span depends also on the digitizer sampling rate.
14
Digitizer Input
IN1
IN2
Band
Center
0 to 625 MHz
312.5 MHz
Maximum Span (Analysis Bandwidth)
625 MHz (650 MHz in interleaved mode)
625 MHz (1.25 GHz in interleaved mode)
M9203A Hardware Extension of 89600 VSA Software
Span
Start
Stop
625 MHz
0
625 MHz

Advertisement