Keysight Technologies M9203A Configuration And Measurement Instructions page 13

High-speed digitizers hardware extension of 89600 vsa software
Table of Contents

Advertisement

Measurement Setup Parameters
This section provides measurement and parameter setup information which are
specific to the 89600 VSA software when used along with the M9203A High Speed
Digitizer. This information can help you to properly setup the VSA in this hardware
configuration to make measurements using your digitizer.
Measurement Setup Parameters:
Alignment / Calibration
Channels
Connection
Coupling
Center Frequency Range Span
Frequency Counter
Hardware
Overlap
Alignment/Calibration: Whenever the digitizer configuration is changed, an internal
calibration is required if this new configuration has not yet been calibrated in the
current session. A manual calibration can be performed at any time from the menu
Utilities > Calibration.
M9203A Hardware Extension of 89600 VSA Software
Preset
Range
Recording
Setup Save/Recall
Trigger Holdoff
Triggering
The 89600 VSA software uses asynchronous polling to check the
calibration status of connected instruments. A
message may appear after a configuration change. This message
disappears after performing a new calibration.
If a signal is present on the input of the digitizer, it may affect the
internal calibration. If the input signal is preventing a successful
internal calibration, a pop-up warning message is raised. In such a
case, please perform a manual calibration with the signal turned
off or disconnected to allow a successful internal calibration .
Measurement Setup Parameters
CAL: Needed
13

Advertisement

Table of Contents
loading

Table of Contents