8. Locate the W45 cable with color code 2 on one end and 17 on the other end. Attach color code
2 end to the Wideband Analog IF at J2. Route the cable as shown in
the cable will be routed later to the A3 Digital IF.
9. Locate the W34 cable with color code 4 at one end and color code 718 at the other end. Attach
the color code 4 end to Wideband Analog IF at J4, and the other end to A16 Reference
assembly at J718.
10.Locate W41 cable with color code 17 at one end and color code 901 at the other end. Attach
the color code 17 end to the Wideband Analog IF at J17, and the other end to A15 Front End
Controller at J901. Remove cable that is on A15J901, it will be discarded after the other end of
the cable at A3J15 is removed later.
11.Locate W40 cable with color code 18 at one end and color code 15 at the other end, Attach the
color code 18 end to the Wideband Analog IF at J18. The other end of the cable will be routed
to the A3 Digital IF assembly.
12.See
Figure 2
that shows the hole in the chassis the W40 cable routes through in order to
connect to A3J15.
Figure 2
Cable Routing
Installation Note N9040-90017
Option B2X, 255 MHz Analysis Bandwidth
Figure
2, the other end of