Find Quality Products Online at:
[How to attach]
F irmly insert and connect the adapter to the end of the cord (with
banana plugs at both ends).
Alligator clip
Flat test bar
# DANGER
To avoid getting electrical shocks, ensure that test leads are
disconnected from the instrument when replacing the metal tip or
adapter for test leads.
6-3 Battery voltage check
(1) P lease refer to 11. Battery replacement in this manual and
insert batteries in KEW3431.
(2) S et the function switch to any position other than OFF to power
on theinstrument.
(3) Check the color of LED battery status indicator.
Solid green : Battery voltage is enough.
Blinking green : Battery voltage is low.
R eplace batteries with reference to 11. Battery replacement
to perform further measurements.
Solid red : Battery voltage is below the lower limit of the operating
v oltage. In such a condition, accuracy of the measured result
isn t guaranteed. Replace the batteries with new ones as soon
as possible.
● C olor of LED battery status indicator might change from green to
red depending on measured objects; for example, resistance of
the measured object is low.
● U se of AA alkaline battery is recommended. Use of other
batteries may cause improper indication of battery level.