8.
3.2.2 Constant voltage operation mode (CV)
Under CV mode, the electronic load will consume sufficient current to maintain
the input voltage at setting voltage. As shown in Fig. 3-2.
Load
Fig. 3-2 Voltage-Current Relation Schema under CV Mode
Under CV mode, the electronic load provides three ways to modify constant
voltage.
Operation steps
1.
2.
3.
Constant
Current
Rise down=0.000A/uS
Complete parameter setting.
10.0000V
0.0000A
0.00W CC=1.000A
NOTE
If the above method is for editing auto test step (as mentioned below), constant current
range can also be set.
V
voltage
Set current
CV mode
Under CV mode, rotate the pulse knob to change constant voltage value.
Under CV mode, directly input numeric key and press [Enter] key for
confirmation to change constant voltage value.
Under CV mode, move the cursor with Left/Right Key and press Left/Right
Key to adjust value at corresponding position.
Press [CV] key and [Shift] + [CV] to enter parameter setting screen.
Constant
Voltage
Range=120.00V
Set maximum working voltage value and press [Enter] key for
confirmation.
Constant
Voltage
Range=2.33V
Set maximum current value and press [Enter] key for confirmation.
Constant
Voltage
High=66.000A
Copyright © ITECH Electronic Co., Ltd.
Function and Features
Set voltage
I
22