Chapter3 Configuration Commands; [Channel:]Voltage:group[:Enable] <Nr1>,<Nr2; [Channel:]Resistance:group[:Enable] <Nr1>,<Nr2 - ITech IT5102 Programming Manual

Hide thumbs Also See for IT5102:
Table of Contents

Advertisement

Chapter3 Configuration Commands

[CHANnel:]VOLTage:GROUp[:ENABle] <NR1>,<NR2>
This command sets the voltage measurement function to enable or disable for
all channels in the corresponding group. The parameter NR1 is the group
number, one instrument is a group, and when multiple instruments cascade, the
slave instrument address is a group number.
NR2 is a decimal number. The 16 bits corresponding to the conversion of NR2
into binary represent the enabling/disabling of the voltage measurement
function of the corresponding channel; 1 represents the disabling, and 0
represents the disabling.
Command syntax
VOLTage:GROUp[:ENABle] <NR1>,<NR2>
Parameter
NR1: Group 0~16
NR2: enable or disable, 0~255(IT5102E), 0~65535(IT5102)
Example
VOLT:GROU 1,7
Set voltage measurement function of channel 1, channel 2 and channel 3 in
group 1 to enable.
Query Syntax
VOLT:GROU?
Return Parameters
<NR1>,<NR2>
Relevant Command
VOLT:GROU?
[CHANnel:]RESistance:GROUp[:ENABle]
<NR1>,<NR2>
This command sets the internal resistance measurement function to enable or
disable for all channels in the corresponding group. The set method is the same
as [CHANnel:]VOLTage:GROUp[:ENABle] <NR1>,<NR2>.
Command syntax
RESistance: [:ENABle] <NR1>,<NR2>
Parameter
NR1: Group number 0~16